Catalogue numbers for GORE® Molding & Occlusion Balloon

| GORE® Molding & Occlusion Balloon | ||||
|---|---|---|---|---|
| Catalogue Number | Balloon Inflation Range (mm) | Balloon Length (mm) | Catheter Length (cm) | Wire Compatibility (in) | 
| MOB37 | 10-37 | 40 | 90 | 0.035 | 
| Inflation volume guide | |
|---|---|
| Intended Balloon Diameter (mm) | Recommended Inflation Volume (ml) | 
| 10 | 2 | 
| 20 | 7 | 
| 30 | 16 | 
| 37* | 35 | 
| *Maximum balloon inflation diameter 
 CAUTION: The above balloon inflation chart volumes and diameters are only an approximate guide. Balloon inflation should only be performed while closely monitoring under fluoroscopy. Do not inflate the balloon to a diameter larger than the vessel it is being inflated within. | |

Refer to Instructions for Use at eifu.goremedical.com for a complete description of all applicable indications, warnings, precautions and contraindications for the markets where this product is available. RXOnly
INDICATIONS FOR USE IN EUROPE: The GORE® Molding and Occlusion Balloon Catheter is intended for temporary occlusion of large diameter vessels or to assist the expansion of self-expanding endovascular prostheses (stent grafts). 
CONTRAINDICATIONS: The GORE® Molding and Occlusion Balloon Catheter is contraindicated in patients who:  
- are contraindicated to contrast media or anticoagulants
- have an arterial entry site that cannot accommodate a 10 Fr introducer sheath
- are minors
- are pregnant.