Freeman J, Chen A, Weinberg RJ, Okada T, Chen C, Lin PH. Sustained thromboresistant bioactivity with reduced intimal hyperplasia of heparin-bonded PTFE Propaten Graft in a chronic canine femoral artery bypass model. Annals of Vascular Surgery 2018;49:295-303.

Abstract

Introduction:

Bypass graft thrombosis remains a significant mode of failure in prosthetic graft revascularization. The purpose of this investigation was to evaluate the long-term thromboresistant effect of heparin-bonded expanded polytetrafluoroethylene (ePTFE) graft using Carmeda BioActive Surface technology in a canine model.

Methods:

Bilateral femorofemoral artery bypass grafts with ePTFE grafts were performed in 25 adult grayhound dogs. In each animal, a heparin-bonded ePTFE graft (Propaten, WL Gore) was placed on one side, whereas a control nonheparin graft was placed on the contralateral side. The graft patency was assessed at 1, 6, 12, 18, and 24 months (n = 5 per group) following the bypass. Heparin bioactivity of the graft material was analyzed. The effect of intimal hyperplasia was also assessed.

Results:

All bypass grafts were patent at 1 month. Significantly greater patency rates were noted in the Propaten group compared to the control group at 12, 18, and 24 months, which were 84%, 80%, and 80% vs. 55%, 35%, and 20%, respectively (P < 0.02). There was a significant reduction in the anastomotic neointimal area and neointimal cell proliferation in Propaten grafts compared with control grafts at all groups between 6 and 24 months (P < 0.05). Heparin bioactivity as measured by antithrombin binding assay was demonstrated in the Propaten graft between 1 and 24 months. Mean heparin activities on Propaten grafts ranged from 26.3 ± 6.4 pmol/cm2 to 18.4 ± 8.7 pmol/cm2 between 1 and 24 months, which were significantly greater than the control group (P < 0.001). Differences between mean heparin activities of explanted Propaten graft samples at the various time points were nonsignificant (P > 0.05).

Conclusions:

Heparin-bonded ePTFE graft provides a thromboresistant surface and reduced anastomotic intimal hyperplasia at 2 years. The stable heparin bioactivity of the Propaten graft confers an advantage in long-term graft patency.


Disclaimer:

Gore products referenced within, if any, are used within their FDA approved/cleared indications. Gore does not have knowledge of the indications and FDA approval/clearance status of non-Gore products. Gore makes no representations as to the surgical techniques, medical conditions, or other factors that may be described in the article(s). The reader is advised to contact the manufacturer for current and accurate information.


IFU Consult instructions

Refer to Instructions for Use at eifu.goremedical.com for a complete description of all applicable indications, warnings, precautions and contraindications for the markets where this product is available. RXOnly

INDICATIONS FOR USE IN EUROPE: GORE® PROPATEN® Vascular Graft is indicated for patients who require replacement or bypass of peripheral vessels due to occlusive or aneurysmal diseases or trauma, or for patients who require arteriovenous hemodialysis access in the upper or lower extremities due to end-stage renal disease (ESRD).

GORE® PROPATEN® Vascular Graft Configured for Pediatric Shunt is indicated for staged palliative repair of cyanotic congenital heart disease (CCHD).

CONTRAINDICATIONS: DO NOT use the GORE® PROPATEN® Vascular Graft in patients with known hypersensitivity to heparin, including those patients who have had a previous incidence of Heparin-Induced Thrombocytopenia (HIT) type II.

DO NOT use any of the below configurations of GORE® PROPATEN® Vascular Graft for coronary artery bypass or cerebral reconstruction procedures:

  • GORE® PROPATEN® Vascular Graft Integrated Rings
  • GORE® PROPATEN® Vascular Graft Fixed Ring
  • GORE® PROPATEN® Vascular Graft Removable Ring
  • GORE® PROPATEN® Vascular Graft Removable Ring Axillobifemoral

DO NOT use GORE® PROPATEN® Vascular Graft as a patch. If cut and used as a patch, GORE® PROPATEN® Vascular Graft may lack adequate transverse strength.

FOR PATCHING APPLICATIONS: For cardiovascular procedures requiring patch materials, use the appropriate GORE® ACUSEAL Cardiovascular Patch.

Product may not be available in all countries. Please check with your Gore representative for availability.